Optical | |||
Minimum | Operation | Max. rating | |
---|---|---|---|
Wavelength [nm] | 531 | 532 | 533 |
CW output peak power [W] | 6 | 8 | |
QCW output peak power [W] | 13 | 15 | |
Output beam size [µm] | 120 x 80 | ||
Beam position (height) [mm] | 22 | ||
Divergence [mrad] | 10 | ||
M2 | 4 | ||
Electrical | |||
LD operating current, typ. [A] | 10 | 28 | 40 |
LD operating voltage [V] | 1.6 | 1.7 | 1.9 |
Monitor PD voltage (@3W, 10 kW) [Vdc] | 1.5 | 2 | |
LD TEC current [A] | 8 | ||
LD TEC voltage [Vdc] | 14 | ||
NLC TEC current [A] | 2 | ||
NLC TEC voltage [Vdc] | 4.8 | ||
Laser pulse duration [ms] | 10 | 2500 | |
Repetition rate [Hz] | 1 | 25 | CW |
Rise and fall time(1) [ms] | 0,5(2) | 2 | 5(3) |
Warm up time(2) [s] | 30 | ||
Power stability, short time (RMS) [%] | 1 | ||
Mechanicss | |||
Operation temperature (housing)(3) [°C | 15 | 45 | |
Weight [g] | 900 | ||
Dimensions [mm3] | 67 x 42 x 167 | ||
Expected Lifetime [h] | 10000 | ||
Laser class (EN-60825) | 4 |
1. Rise time for pulsed conditions, measured at 25 Hz and 50% DC
2. Defined as the time required for laser stabilization (time to establish LD and NLC operation temperature and crystal thermal lens). In pulsed mode, the warm up time only affects the first pulses.
3. Within the temperature range, optical power could vary ±10%, to be compensated from the monitor photodiode signal through PID.
For higher temperatures, laser head should be mounted on a cooled surface with a capacity to remove 60 W waste heat at max. housing
This content cannot be seen on mobile view. Change to PC or tablet landscape view.
Possible levels of integration for this product are marked in red
MICRO-OPTICS
MICRO-OPTICS
HOUSING
PRISM
CRISTALS
MICRO-OPTICS
HOUSING
PRISM
CRISTALS
ELECTRONIC COMPONENTS
FIBER OUTPUT
Dermatology & Aesthetics |
Diode lasers with different level of integration, from diode laser stacks up to plug & play laser modules. More info |
Material Processing |
High power lasers for use in material manufacturing as Welding, Drilling, Cutting, Cladding, Engraving or Cleaning More info |
Ophtalmology |
Low power, single frequency diode lasers and high energy diode-pumped solid state laser for coherent, remote and common Raman spectroscopy set-ups. More info |
Optical Pumping |
Pumping sources involving single emitters, emitter arrays and vertical and horizontal laser bar mounts and stacks More info |
We develop your laser solutions, to your needs. Fill in the form to define your desired parameters.